Operator’s Manual – OPTI CCA-TS2 8-13
8 DIAGNOSTICS AND TROUBLESHOOTING
<Not Ready: Gas Not Installed>
The gas bottle is empty or has not been installed
properly. If you remove a gas bottle that is still
valid, you will have to reinstall the same gas bottle.
• Press <New Gas Bottle> and reinstall the gas
bottle (See Section 7.5.1).
<Not Ready: Gas Expired>
The in-use (6 months) or labeled shelf-life of the gas
bottle has expired.
• Press <New Gas Bottle> and replace the gas
bottle (see Section 7.5.1).
<Not Ready: Low Battery>
The battery voltage is low.
• Operate the analyzer on AC power and/or
recharge the battery.