Operator’s Manual – OPTI CCA-TS2 8-23
8 DIAGNOSTICS AND TROUBLESHOOTING
8.2.10 Checking the Factory Settings (FSet)
The <FSet> function (Fig. 8-11) is designed
exclusively for use by authorized OPTI Medical
personnel and requires a special User ID and
password.
• Press
to return to the <System> screen
or
to return to the main menu.
Fig. 8-11 Factory Settings
8.2.11 Checking the Bar Code Scanner
This option allows you to check the function of the
bar code scanner.
In the main menu, press <System Manager ->
Diagnostics>.
• Select the <Tests> tab.
Press <Barcode> to start the test (Fig. 8-12).
• To test the bar code scanner, scan a bar code
label of e.g. a sensor cassette (Fig. 8-13).
The display will show a sequence of numbers
(Fig. 8-14). Compare the numbers with those
printed on the cassette bar code label. Matching
information conrms the proper function of the
bar code scanner.
• Press
to return to the <Tests>
screen.
• Press
to return to the <System> screen
or
to return to the main menu.
Fig. 8-12 Barcode Test
Fig. 8-13 Scan Barcode
Fig. 8-14 Barcode Test