EasyManua.ls Logo

OPTi CCA-TS2 - As Needed Maintenance; Changing the Gas Bottle

OPTi CCA-TS2
299 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator’s Manual – OPTI CCA-TS2 7-9
7 MAINTENANCE
The calibration gas bottle is designed to provide
approximately 80 sampling operations (45 for
B-Lac). The following message will alert the
operator that the gas bottle needs to be changed
(Fig. 7-21).
To change the gas bottle:
1. Press <New Gas Bottle>.
2. Unscrew the gas bottle by turning the knob on
the bottom counterclockwise.
3. Take a new gas bottle and remove its cap.
4. When prompted (Fig. 7-22), scan the new gas
bottle bar code on the insert sheet by holding it
2-3 inches (5-8 cm) from the bar code scanner
on the bottom right-hand corner of the analyzer.
• The red line from the barcode scanner
should cover the entire barcode.
• The analyzer will beep when the barcode is
accepted.
• Record the date of installation on the gas
bottle for later reference.
NOTE: Iftheinsertsheetismisplaced,youcan
enterthelotnumberonthegasbottlelabel
manually.Press<Manual>(Fig.7-22)and
enterthenumberusingthenumerickeypad.
NOTE: Thebarcodecontainsexpiration
information.TheOPTICCA-TS2willalert
theoperatorfourweeksbeforethegasbottle
expires.
NOTE: Thegasbottleshouldalwaysbestoredwith
thecapon.
5. When prompted (Fig. 7-23), install the new gas
bottle.
6. Insert the bottle into its housing and turn it
clockwise until nger-tight (Fig. 7-24).
Press (Fig. 7-23).
7.5 As Needed Maintenance
7.5.1 Changing the Gas Bottle
Fig. 7-21 Select New Gas Bottle
Fig. 7-24 Gas Bottle
Fig. 7-22 Scan Bar Code
Fig. 7-23 Insert Gas Bottle

Table of Contents