EasyManua.ls Logo

OPTi CCA-TS2 - Checking the Pump Flow

OPTi CCA-TS2
299 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator’s Manual – OPTI CCA-TS2 8-27
8 DIAGNOSTICS AND TROUBLESHOOTING
8.2.15 Checking the Pump Flow
This option is designed to test the pump cartridge.
In the main menu, press <System Manager ->
Diagnostics>.
• Select the <Tests> tab.
Press <Flow> to start the test (Fig. 8-22).
• Insert a new cassette (Fig 8-23).
• Close the SMC cover.
• Wait for test results (Fig. 8-24).
• The two numbers indicate the actual ow rates
clockwise and counter clockwise (Fig. 8-25).
If one of the two or both rates are out of range,
the test fails.
• Repeat test or replace the pump cartridge, if
the test fails. See replacement instructions in
Chapter 7.4.1.
NOTE: Itispossiblethatthetestfailstherst
time,evenifthepumpcartridgeisworking
correctly.
• Remove the cassette (Fig. 8-26).
Fig. 8-22 Flow Test
Fig. 8-24 Flow Test in Process
Fig. 8-23 Insert Cassette
Fig. 8-25 Flow Test Pass
Fig. 8-26 Remove Cassette

Table of Contents