Operator’s Manual – OPTI CCA-TS2 3-29
3 CUSTOMIZATION
Fig. 3-67 Printer Conguration
The OPTI CCA-TS2 can be connected to an external
printer. In the <Conguration> menu (Fig. 3-67),
you can select the printer conguration.
1. Select the options to be enabled.
2. Press to accept the changes.
3. Press to return to the <Setup> screen or
to return to the main menu.
3.2.4.2 Dening Units
This menu lets you change the units of measure
for pressure, temperature, output resolution, total
hemoglobin, Ca
++
, Glu, BUN (urea) and Lac.
1. In the main menu, select <System Manager>
and <Setup>.
2. In the <System Setup> menu, press the
<System> tab, then press <Units> (Fig. 3-68).
3. In the <System> screen, select the units for the
displayed parameters (Fig. 3-69).
4. The selection for <Resolution> on this menu
determines the number of digits displayed and
printed past the decimal point, for all measured
parameters.
NOTE: Theselectionappliestopatientsample
resultsonly.Bydefault,theresolutionfor
patientsamplesishigh.
ResolutionisalwayshighforControland
SRCresults.
Fig. 3-69 System
Fig. 3-68 Select Units