9-6 Operator’s Manual – OPTI CCA-TS2
9 OPERATING PRINCIPLES
9.5.2 Test Conditions
Sample Size: a minimum of 125 µL (60µL for B60 cassette)
Sample Type: heparinized whole blood, serum and plasma
Sample Application: syringe, capillary or ComfortSampler
Ambient Temperature: 10 - 30 ºC (50 – 86 ºF)
Relative Humidity: 5% to 95% (non-condensing)
Type of Measurement: optical uorescence (pH, PO
2
, PCO
2
, Na
+
, K
+
, Ca
++
) Cl
-
,
Glu, BUN (urea), Lac) and reectance (tHb, SO
2
)
9.5.3 Input Values
Parameter Ranges/Options Default
Patient ID 25 alphanumeric characters Blank
Accession Number 25 alphanumeric characters Blank
Date of Birth Month, DD, YYYY
Patient Sex Male, female or unknown unknown
Patient temperature, T 14.0 to 44.0 °C 37.0 °C
57.2 to 111.2 °F 98.6 °F
Medical Record Number 25 alphanumeric characters Blank
Account Number 25 alphanumeric characters Blank
Test ID 25 alphanumeric characters Blank
Patient Name 25 alphanumeric characters Blank
Age 1-150 0
Attending Physician 25 alphanumeric characters Blank
Patient Location 25 alphanumeric characters Blank
Sample Collection Time Month, DD, YY, HH:MM Blank
Sample Type Art, Ven, MixVen, Cap, Art
Cord, CPB, where:
Art = Arterial
Ven = Venous
MixVen = Mixed Venous
Cap = Capillary
Cord = Cord
CPB = Cardio-Pulmonary Bypass