Glu-6 Operator’s Manual – OPTI CCA-TS2
GLUCOSE ANALYTES
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
Parameter and Specication Value
Number of Samples 54
Range 69.4 to 361.3
Slope 0.99
Offset -0.44
R
2
0.983
Sy|x 12.849
References
1. Tietz, Norbert W., Ed., ClinicalGuidetoLaboratoryTests, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. OPTI Medical. Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b).
50
100
150
200
250
300
350
400
50 100 150 200 250 300 350 400
Linear fit
(-0.44 + 0.99x)