3-16 Operator’s Manual – OPTI CCA-TS2
3 CUSTOMIZATION
• To change units of measure, see section 3.2.4.2.
NOTE: Eachfacilityshouldestablishitsown
referenceranges.Thepresetanalyzerranges
areforreferenceonlyandarederivedfrom
“Tietz,BurtisC,etal(Eds.),Textbook
ofClinicalChemistryandMolecular
Diagnostics,4thEd.,(ElsevierSaunders,
2006)pps.2252-2302”.
5. Select the <Critical> tab (Fig. 3-34).
By default, critical ranges are disabled.
Press <Enable> to enter critical limit values.
6. Select the parameter you want to change and
press to enter the new value.
7. Press to access pages 2, 3 and 4 with
the remaining parameters (Figs. 3-35 and 3-36)
and press
to accept the new values.
The instrument is preset to the measurement ranges
of the TS2:
HCO3
-
: 1.0 - 200.0 mmol/L
BE: -40.0 - +40.0 mmol/L
BE
ecf
: -40.0 - +40.0 mmol/L
BE
act
: -40.0 - +40.0 mmol/L
BB: 0.0 - 100.0 mmol/L
tCO2: 1.0 - 200.0 mmol/L
st.HCO3
-
: 1.0 - 200.0 mmol/L
st.pH: 6.500 - 8.000
SO2(c): 0.0 - 100.0 %
O2ct: 0.0 - 56.0 mL/dL
Hct(c): 15.0 - 75.0 %
cH
+
: 10.0 - 1000.0 nmol/L
AaDO2: 0.0 - 800.0 mmHg
AnGap: 3.0 - 30.0 mmol/L
P50: 15.0 - 35.0 mmHg
nCa
++
: 0.10 - 3.00 mmol/L
• In all data input screens, if values outside the
acceptable input range are entered, the system
automatically ags the error and displays the
valid range.
•
The limits entered here will reside in the
instrument memory even after system power is
turned off.
8. Press to return to the <Setup> screen or
to return to the main menu.
Fig. 3-34 Enter Critical Limits
Fig. 3-35 Critical Limits 2