EasyManua.ls Logo

OPTi CCA-TS2 - Page 75

OPTi CCA-TS2
299 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator’s Manual – OPTI CCA-TS2 3-39
3 CUSTOMIZATION
4. Go back to the <System>Setup>Ethernet>
screen (Fig. 3-88) and ll in the following elds
as needed:
<IP Type> – choose static or dynamic IP Type.
<IP Address> – static IP address of the
instrument. Enter the IP address, if IP type is
set to static. This address can be found in the IP
address eld in the <Ethernet Test> screen
(Fig. 3-87 above).
<Subnet Mask> – Enter the subnet mask,
if IP type is set to static. The subnet mask is
specic to your network. Ask your network
administrator if you are unsure of the value.
<Gateway> – Enter the gateway address, if
IP type is set to static. The gateway address
is network specic. Ask your network
administrator if you are unsure of the value.
<Discovery Port> – is used for instrument
discovery. Use for static and dynamic IP types.
The discovery port is set to 6703 by default, but
can be changed if needed.
NOTE:IfstaticIPtypeisselected,theinstrument
IPaddressandGatewayneedtobeonthe
samenetworktocommunicate.
5. Press to accept the changes.
6. Press to return to the <Setup> screen or
to return to the main menu.
When network connection is enabled, the status bar
displays the network icon
(Fig. 3-88).
Fig. 3-88 Ethernet Settings

Table of Contents