Page 3-6 HandiLaz Mini Operator’s Manual
Chapter 3: MEASURE MODE Settings
Setting CALC (Calculation) Sample Parameters
The parameters seen when the CALC measure mode is opened are those that
were saved when the CALC mode was last run. The CALC measuring mode can
be run as it is, without changing any sample parameters.
To edit the setup screen:
1
Press the
ENTER
key to move the cursor in a loop through each field that can be
edited.
2
At the field to be changed, press the up or down key.
The field setting will loop through its range of allowable settings.
3
When the field is set correctly, press the
ENTER
key again to move to the next
field that can be edited.
To close the edit function and run the sample:
1
Use the
ENTER
key to navigate to NO.
2
Use the up or down key to toggle the NO to OK.
3
Press the
ENTER
key.
The SAMPLING display will appear.
Taking CALC (Calculation) Sample
From the SAMPLING display, press the
ENTER
key. After the initial delay,
press
STOP/START
to begin sampling.
ISO Sampling Mode
Use this mode to sample and perform ISO calculations. Because of the greater
time required to sample in ultraclean rooms, this 0.1 CFM sample flow
instrument is the most time-efficient in certifying rooms of ISO Class 5 through
8. The ISO measurement mode considers only 0.5 and 5.0 µm particles.
To open the ISO sampling mode:
1
From the main menu, use either arrow key to navigate to MEASURE MODE.
2
Press the
ENTER
key.
The MEASURE MODE menu will open.
SAMPLING
SAMPLE TIME 00:00:45
TOTAL TIME 00:01:15
FREE REC.:09500 OK