HandiLaz Mini Operator’s Manual Page D-1
Appendix D
Error Messages and Alarms
When sampling, HandiLaz Mini monitors its operation, available power, and
alarm settings. When an error or warning condition occurs, the display changes
to alert the user. Indications of an error appear in the following forms:
An Error code blinks in the form of blinking letters (L, F, O, or M) to the left
of the time display.
A low battery power symbol appears to the right of the time display.
The particle size label(s) blinks to indicate that a particle limit setting has
been exceeded.
Error Code and Alarm Locations
REPEAT 3B
WAIT
0.3
0.5
M
1,507 CNT
623 CNT
15:25
Blinking indicates that a particle
alarm limit has been exceeded.
Low battery
power icon