Saving Data to Other MediaSaving Data to Other Media
LASAIR II Operator’s ManualLASAIR II Operator’s Manual
PagePage
6-176-17
Saving Data to Other MediaSaving Data to Other Media
The LASAIR II saves 3,000 The LASAIR II saves 3,000 sample data records internally. Once the record filesample data records internally. Once the record file
is full, a new record will displace the oldest record is full, a new record will displace the oldest record in the file. A LASAIR II setin the file. A LASAIR II set
to sample once a minute will fill the record to sample once a minute will fill the record file in about 48 hours.file in about 48 hours.
If you want to save the If you want to save the sample data, monitor the status of your LASAIR II’s datasample data, monitor the status of your LASAIR II’s data
file and copy the sample datafile and copy the sample data
beforebefore
the data file is completely the data file is completely full. Remember,full. Remember,
once the data file is full, youonce the data file is full, you
willwill
lose the oldest data when new sample data islose the oldest data when new sample data is
collected.collected.
Monitoring the Sample Data FileMonitoring the Sample Data File
To monitor the status of the sample data file from the LASAIR II:To monitor the status of the sample data file from the LASAIR II:
11 Press the Information Icon softkey to open the Information Screen.Press the Information Icon softkey to open the Information Screen.
22
Observe the condition of the Data: “thermometer indicator”. The Observe the condition of the Data: “thermometer indicator”. The indicator willindicator will
display the number of records stored and adisplay the number of records stored and as a relative degree to which thes a relative degree to which the
thermometer is filled. The white space in the bar thermometer is filled. The white space in the bar is a relative indication of theis a relative indication of the
amount of space remaining in the amount of space remaining in the data file.data file.
Saving the Sample Data FileSaving the Sample Data File
YYou may want to ou may want to copy the sample dacopy the sample data to another file so it can ta to another file so it can be saved for long-be saved for long-
term retention or for data analysis. This can be term retention or for data analysis. This can be done by any of done by any of the followingthe following
methods:methods:
Data can be downloaded in CSV format by acData can be downloaded in CSV format by accessing the LASAIR II from acessing the LASAIR II from a
web browser/Ethernet connection and using the LASAIR web browser/Ethernet connection and using the LASAIR II data saveII data save
featufeature. See Chaptre. See Chapterer 7 and Appendi7 and Appendix E.x E.
Data can be saved cData can be saved continuously by connecting your LASAIR II to ontinuously by connecting your LASAIR II to a facilitya facility
management system such as Facility Net or Pharmaceutical Net andmanagement system such as Facility Net or Pharmaceutical Net and
collecting your samples while so connected. When using such collecting your samples while so connected. When using such a system, thea system, the
data is saved continuously on a PC, and the data is saved continuously on a PC, and the local data storage can be ignored.local data storage can be ignored.
See Appendix B and C.See Appendix B and C.
Data can be saved from a Data can be saved from a mobile LASAIR II by periodically connecting tomobile LASAIR II by periodically connecting to
Facility Net to download the LASAIR II’s sample data. See “PreparingFacility Net to download the LASAIR II’s sample data. See “Preparing
LASAIR II for LASAIR II for use in Remote Mode” on pageuse in Remote Mode” on page 7-7. Also see Appendix C.7-7. Also see Appendix C.
Data can be downloaded in CSV format by means of the USB port. SeeData can be downloaded in CSV format by means of the USB port. See
Appendix J.Appendix J.

Table of Contents

Other manuals for Particle Measuring Systems LASAIR II 550L

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Particle Measuring Systems LASAIR II 550L and is the answer not in the manual?

Particle Measuring Systems LASAIR II 550L Specifications

General IconGeneral
BrandParticle Measuring Systems
ModelLASAIR II 550L
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals