Appendix-A Error codes explained
NO PROPER CALIBRATION AVAILABLE. (No calibration points available).
UNDERFLOW. (A/D reads al 0’s, check loadcell)
OVERFLOW. ( A/D reads al 1’s, check loadcell )
DISPLAY OVERFLOW. (Exceed maximum display value) (IND-1)
Weigher error code:
2001 - Parameter error
2005 - Input value is not valid
2101 - Weigher not stable
2111 - Flash ROM exhausted
2112 - Error on header creation
2103 - Parameter below zero
2113 - Error on data write
2104 - Not in zero range
2114 - Header validation failed
2105 - Arithmetic overflow occurred
2115 - De-active old data fail
2106 - A/D reads all 1’s
2107 - A/D reads all 0’s
2109 - Gain > 0.999984741211
2102 - Parameter exceeds maxload
2108 - Gain ref.< Zero ref.
2110 - Save error
2117 - Item not found in store
2118 - Error in stored data
2119 - Bad calibration
2120 - Action not enabled
2116 - Load errors
2603 - No recipes available
2601 - No program available