8) Saving and Transferring Parameter Settings
The main unit parameter settings can be saved to a memory card, and the parameter
settings on a memory card can be transferred to the main unit.
※ The file name on the memory card is "PARA" and the extension is ".bin".
※ If the file name and extension are changed to anything other than "PARA.bin",
the logger will be unable to recognize the file.
※ Do not format the memory card. A formatted memory card cannot be used.
※ If the setting was ended with “PP--2” by pressing the CLEAR key for 3 seconds,
it is possible to operate the unit using the parameters saved on the memory card.
However, ordinarily, end with “PP--1”, and operate using the parameters in the
main unit memory “PP--1".
-49-
<Saving the parameters in the main unit to the memory card>
1. Verify that [READY] is displayed on the display screen. If [READY] is not
displayed, press the CLEAR key to display it.
2. Insert the memory card into the main unit.
3. Press and hold switch 1, the SHIFT key, and the MODE key at the same time for
3 seconds. The screen changes to the parameter selection screen, and either
PP--1 or PP--2 is displayed.
4. Press the SET key and display PP--1.
5. Press and hold the SHIFT key for 3 seconds. The display turns off for an instant,
and the parameters in the main unit memory are saved onto the memory card.
6.
Verify that PP--1 is displayed, then press and hold the CLEAR key for 3 seconds.
The screen returns to the initial screen, and [READY] is displayed.
READY
To parameter position screen
+
SHIFT