Appendices . 93
Initialization errors
If your instrument generates one of the following error messages, switch off the instrument,
then switch the instrument on again.
Consult your PerkinElmer service representative if the problem persists.
Error message Action
Program Integrity
Failed
The instrument’s firmware may be corrupt, so must be reinstalled.
Instrument
Configuration Failed
Suggests that one or more internal device is failing. If the problem
persists, note any associated error messages and consult your
PerkinElmer service representative.
Scan Failed If the problem persists, note any associated error messages and
consult your PerkinElmer service representative.
Laser Failed
Replace Laser
Refer to
Replacing the Laser and Power Supply
on page 84.
Source Failed
Replace Source
Refer to
Replacing
on page 74.
Calibration Failed Too little energy may be reaching the detector. Make sure the
sample compartment is clear.