INSTALLING THE ASSISTANT ELEMENT
124 Planmeca Sovereign
Installation manual
Carefully lift the assistant element to its position. Make sure that the attachment plate is
positioned narrower end foremost. Tighten the two attachment nuts using a 13 mm fork
spanner.
Route the cables and tubes into the cuspidor base and connect them as shown on the figure
below.
S37_5 .eps
01-J14
01-J7
Suction
Ground cable
USB
01-J14
01-J7
Suction
Air 6/4 Black
Rinsing water6/4 Green
Instr. water 3.2/1.6 Green
Return water 3.2/1.6 Red
Ground cable
Air 6/4 Black
Return water
3.2/1.6 Red
Instr. water
3.2/1.6 Green
Rinsing water
6/4 Green