Factory presets for Tooth, Teeth, Jaw, Face, Skull, Ear, Temporal Bone and Vertebrae
programs
Child (XS) Small adult (S) Medium-sized
adult (M)
Large adult (L) Extra-large adult
(XL)
Voxel
size
kV mAs kV mAs kV mAs kV mAs kV mAs
600 90 14 90 16 100 12.5 110 12.5 120 11
450 90 18 90 20 100 16 110 16 120 14
300 90 22 90 25 100 20 110 20 120 18
200 90 28 90 32 100 25 110 25 120 22
150 90 36 90 45 100 32 110 32 120 28
75 90 63 90 71 100 50 110 50 120 45
ULD
600 90 7.1 90 8.0 100 6.3 110 6.3 120 6.3
450 90 9.0 90 8.0 100 8.0 110 7.1 120 6.3
300 90 11 90 12.5 100 10 110 9.0 120 8.0
200 90 14 90 16 100 12.5 110 11 120 10
150 90 18 90 20 100 16 110 14 120 12.5
75 90 32 90 32 100 25 110 22 120 20
Factory presets for Sinus, Nose and Airways programs
Child (XS) Small adult (S) Medium-sized
adult (M)
Large adult (L) Extra-large adult
(XL)
Voxel
size
kV mAs kV mAs kV mAs kV mAs kV mAs
600 90 14 90 16 100 12.5 110 12.5 120 11
450 90 14 90 16 100 12.5 110 12.5 120 11
300 90 18 90 20 100 16 110 16 120 14
150 90 28 90 32 100 25 110 25 120 22
ULD
600 90 7.1 90 8.0 100 6.3 110 6.3 120 6.3
450 90 7.1 90 8.0 100 6.3 110 6.3 120 6.3
300 90 9.0 90 10 100 8.0 110 7.1 120 6.3
150 90 14 90 16 100 12.5 110 11 120 10
Follow these steps if you need to adjust the preset exposure values for
current exposure:
Steps
1. Select this field (1) to open a pop-up window.
2. Use the minus or plus signs (2) to set the exposure values you wish to
use. To improve the image contrast, reduce the kV value. To reduce the
radiation dose, reduce the mAs value.