13.1.4 Testing Routines (1500)
13.1.4.1 2D test exposure
About this task
Follow the steps below to take a 2D test exposure.
Steps
1. Select User > 1500 Testing Routines > 1510 2D Test Exposure.
2. Use the - or + button to set the exposure values you wish to use.
3. Select the SET button.
4. Move to a protected area.
5. Press and hold down the exposure button for the duration of the
exposure. The C-arm does not move when you take a test exposure.
6. Select the green check mark button.
13.1.4.2 3D test exposure
About this task
Follow the steps below to take a 3D test exposure.
Steps
1. Select User > 1500 Testing Routines > 1520 3D Test Exposure.
2. Use the - or + button to set the exposure values you wish to use.
3. Select the SET button.
4. Move to a protected area.