Cold Hot
Codes What It Means What to Do
Error 5
Dirty
Optics
E-5
Dust or dirt
in meter.
Remove
cartridge and
gently clean
optics with so,
dry lint-free
cloth. (see Meter
Optics Care on
page 69).
Try testing with
a new cartridge.
Error 6
Barcode
Error
E-6
Meter could not
read barcode data
from cartridge.
Re-insert
cartridge.
If error repeats,
insert
new cartridge.
Error 1
System
Failure
E-1
A problem with
the meter has
occurred.
Remove batteries,
press and hold the
power button for
at least 3 seconds.
Reinsert batteries.
Temperature
out of Range
Alert
Meter
temperature is
out of operating
temperature
range.
Ensure meter
and cartridge
are both at room
temperature
before testing.
76
Customer Support 1-855-IMI-POGO (464-7646) • www.presspogo.com
If other situations occur such as meter fails to turn on aer batteries are replaced,
meter is unable to clear error code or the cartridge cannot be removed, please call
Intuity Medical Customer Support at 1-855-IMI-POGO (464-7646).