H - I - J PROGRAMMER’S MANUAL V8.94.11-
- 81 -
Regulation on pressure using a 4 to 20 mA pressure remote transmitter
Let the pressure remote transmitter be of measuring range of 0 to 10 bar, the pressure to be kept should be adjustable
between 4 and 8 bar.
The pressure to be kept should be 6 bar, reaching the pressure should last at least 30 seconds after start
(for taking care of the supplying unit e g. the well)
The pump starts to convey from 20 Hz on, cannot be driven above 50 Hz, it cannot be operated more than 1 minute at
the minimum frequency (it must be stopped), but in the case of 0.5 bar pressure drop it must be started again.
The reference signal of the regulation has to be given from the terminal
Remark: according to the remote transmitter data, 10 bar corresponds to 100%
source of the regulation ref. signal
regulation reference signal maximum
regulation reference signal minimum
regulation reference signal acceleration
regulation start / stop time limitation
regulation start error signal
regulation start / stop hysteresis
source of the feedback signal
analogue input 1. current upper level
analogue input 1. current lower level
regulation reference signal
PID 1 Ap (proportional gain)
PID 1 Ti (integration time)
PID 1 Ad (diff. element overdriving factor)
At simple pressure regulation the most suitable regulation type is "PI" or "I".
The value of Ad is 0.00. By this, the effect of the differentiating elements (Ad, Td ) is switched off.
Setting of the PID parameters depends on the system, they always have to be adjusted to the actual requirements
The start/stop commands can be given out from the terminal blocks, the control reference signal modification from the
terminal.