Operation 150821/A4 Page 57 of 110
CORRECTION: OPEN
Figure 2-7 illustrates the connection of the Kelvin Clip test leads for an open correction. The
MAIN INDEX display menus corresponding to the OPEN operation are illustrated below.
CORRECTION
< MAIN INDEX >
COMPARE
BINNING
Enter Compare function
Enter Binning function
Enter Correction function
F2
F3
F1
< MAIN INDEX CORRECT >
OPEN
SHORT
Perform SHORT correction
Perform OPEN correction
F3
F1
F2
F4
TRIGGER
< MAIN INDEX CORRECT OPEN >
SINGLE
MULTI
ABORT
Single Frequency test
Exit Correction Function
Multiple Frequency test
F3
F1
F2
F4
< MAIN INDEX CORRECT OPEN >
SINGLE
MULTI
ABORT
F3
F1
F2
F4
OPEN CIRCUIT TEST LEADS
THEN PRESS TRIGGER.
< MAIN INDEX CORRECT OPEN >
SINGLE
MULTI
ABORT
F3
F1
F2
F4
FREQ: 100.0 kHz
Cp : 0.023pF
RANGE: 4
CALIB: 3
MEASURED
PASS ... 100%
PRESS ANY KEY TO ESCAPE
MAIN
INDEX
Initiate OPEN correction
Return to MAIN INDEX display