Page 58 of 110 150821/A4 Operation
CORRECTION: SHORT
Figure 2-8 illustrates the connection of the Kelvin Clip test leads for a short correction. The
MAIN INDEX display menus corresponding to the SHORT operation are illustrated below.
1730T LCR Meter
L
CUR
H
CUR
H
POT
L
POT
QuadTech
(-) (+)
F1
F4
F3
F2
01
l
SHORT
L
CUR
L
POT
H
POT
H
CUR
(+)(-)
QuadTech 1700-03 Test Leads
<MAIN INDEX CORRECT SHORT>
ABORT
MULTI
SINGLE
Cp : 0.023 pF
FREQ.: 100.0kHz RANGE : 4
CALIB : 3
MEASURED PASS ... 100%
PRESS ANY KEY TO ESCAPE
Figure 2-8: SHORT Configuration using Kelvin Clip Test Leads
To access the SHORT correction operation, press [MAIN INDEX] then [F3] = CORRECTION.
CORRECTION
< MAIN INDEX >
COMPARE
BINNING
Enter Compare function
Enter Binning function
Enter Correction function
F2
F3
F1
< MAIN INDEX CORRECT >
OPEN
SHORT
Perform SHORT correction
Perform OPEN correction
F3
F1
F2
F4