Page 64 of 110 150821/A4 Operation
Figure 2-13 illustrates the two 7000 Series BNC to BNC test cable lead sets available for the
1730T LCR Meter. The 7000-01 is one meter in length and the 7000-02 is two meters in length.
The 1730T instrument permits the input of the test cable length for better measurement accuracy.
Refer to paragraph 2.3.9, Cable Length for programming instruction. As always, each time the
test cable is changed the 1730T instrument should be re-zeroed.
DUT
1730T LCR Meter
L
CUR
H
CUR
H
POT
L
POT
QuadTech
(-) (+)
F1
F3
F2
MEAS
DISPLAY
F4
TRIGGER
MAIN
INDEX
SYSTEM
SETUP
01
l
F1
F4
F3
F2
<MEAS DISPLAY>
NEXT PAGE 1/4
PARA : Cs - D
LEVEL : 1.00 V
FREQ. : 100 kHz
Cs : 1.2345 pF
D : 1.2345
+
-
IHPHIL PL
7000-01 is 1 meter in length
7000-02 is 2 meters in length
IH
PH
PL
IL
Figure 2-13: 7000-01 & -02 Cables for the 1730 LCR Meter