- 14 -
Load the weighing pan steadily
avoiding mechanical shocks.
Place weighed loads centrally on
the weighing pan (eccentricity
errors are specified by EN 45501
standard, point 3.6.2.).
Do not apply concentrated force
(total load in one point).
Avoid side loading, in particular
side shocks.
11.2. Zeroing
To zero mass indication press
key. Zero indication and the following
pictograms are displayed:
and . The instrument can be zeroed only
when the indication is stable.
Indication can be zeroed only within ±2% range of maximum
capacity. If the zeroed value is greater than ±2% of the
maximum capacity, then the software displays message
<Err2>, and short sound signal is heard.