EasyManua.ls Logo

RADWAG WTC 2000

RADWAG WTC 2000
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 19 -
Remove the load from the weighing pan and press key.
Mass of an empty weighing pan is determined, this is signalled with
display of 'dash', < - >. Next, text <LoAd> (load weight) and mass
value that is to be loaded, e.g. 100 g, are displayed.
Load the weighing pan with weight of specified mass value and press
key.
Weight mass is determined, this is signalled with display of 'dash',
< - >. Next, text <UnLoAd> (remove weight) is displayed.
Remove the load from the weighing pan, <1.2.CA-u> submenu is
displayed.
12.3. Adjustment Report
Adjustment report is automatically printed (using scale-connected printer) at
the end of each adjustment process. To declare report content go to
<P6.1.CrEP> submenu. For detailed information concerning report content
read later sections of this manual.
13. SCALE PARAMETERS
Scale parameters are set to adjust the weighing device to ambient conditions
(filters) or individual needs (autozero on/off, tare values memory). These
parameters are to be found in <P2.rEAd> submenu. <P2.rEAd> submenu
comprises functions allowing you to adjust your weighing device to ambient
conditions of given workstation.
13.1. Filter
Enter <P2.rEAd / 2.1.FiL> submenu.
Press
key, filter values are displayed successively one by one:
1 - Fast, 2 - Average, 3 – Slow.
Set respective value and press
key to confirm, next go to the home
screen.
The higher filter value, the longer the weighing takes.

Table of Contents

Other manuals for RADWAG WTC 2000

Related product manuals