- 40 -
23.2. Reference Sample Mass Determined by Weighing
Enter <3.4.dEu / 3.4.2.UUt> submenu, set <S_S> value.
Enter <dEu> working mode (Percent weighing), first, text <LoAd> is
displayed for 1 second, then the weighing window.
Load the weighing pan with the reference sample. When the indication is
stable (
pictogram is displayed), press key to confirm the mass.
Mass of the weighed load is automatically set as reference sample mass,
the home screen is displayed along with 100.000% value.
23.3. Reference Sample Mass Determined by Entering the Mass Value
Enter <3.4.dEu / 3.4.2.UUt> submenu, set <Suu> value.
Enter <dEu> working mode (Percent weighing).
Text <SEt_Ut> is displayed for 1 s, next, window for declaring mass of
reference sample.
Enter respective value and press
key to confirm. The home screen is
displayed automatically with 0.000% value.
If the value of entered reference sample mass is greater than
max capacity value, then message <Err Hi> is displayed.
24. WORKING MODE – PEAK HOLD
Peak Hold is a working mode allowing you to snap value of maximum force
applied to the weighing pan during one weighing process.
24.1. Local Settings
To go to local settings enter <3.5.toP> submenu.
3.5.1.Acc
Working mode accessibility
For detailed description read section 19.2.1.
3.5.2.Lo
Lo threshold
For detailed description read section 19.2.4.
24.2. Peak Hold Operation
Enter <3.5.toP / 3.5.2.Lo> submenu, set <Lo> parameter value (Lo
threshold) after exceeding of which maximum force is to be registered.