- 20 -
13.2. Value Release
Enter this parameter to adjust rate of stabilisation of the measurement result.
Depending on the selected option, weighing time is either shorter or longer.
Procedure:
Enter <P2.rEAd / 2.2.APPr> submenu.
Press
key, available values are displayed successively one by one:
F_P - fast and reliable, PrEc – reliable, FASt – fast.
Press
key to confirm, next go to the home screen.
13.3. Ambient Conditions
Parameter relating to ambient and environmental conditions of the workstation.
Enter this parameter and set 'nStAb' value if the ambient conditions are
unfavourable (air drafts, vibrations).
Procedure:
Enter <P2.rEAd / 2.3.Enut> submenu.
Press
key, parameter values are displayed successively one by
one: nStAb – unstable, StAb – stable.
Press
key to confirm, next go to the home screen.
13.4. Autozero Function
'Autozero' function has been designed to enable automatic control and
correction of zero indication. This guarantees precise weighing results.
There are, however, some cases when this function can be a disturbing factor
for the measuring process, e.g. very slow placing of a load on the weighing pan
(load adding, e.g. pouring, filling). In such case, it is recommended to disable
the function.
Procedure:
Enter <P2.rEAd / 2.4.Aut> submenu.
Press
key, parameter values are displayed successively one by one:
YES – autozero function enabled, no – autozero function disabled.