- 51 -
27.5. Manual Printout / Automatic Printout
It is possible to generate printouts both manually and automatically.
• Manual printout is generated for stable weighing result. Load the platform,
wait for a stable result and press
key.
• Automatic printout is generated for stable weighing result. Load the
platform, wait for a stable result. No key needs to be pressed.
Option of temporary weighing results printout is disabled for
verified scale.
Format:
1 2 3 4 -12 13 14 15 16 17 18
stability
marker
space character mass space unit CR LF
Stability marker
[space] if measurement result stable.
[?] if measurement result unstable.
[^] if high limit is out of range.
[v] if low limit is out of range.
Character
[space] for positive values.
[-] for negative values.
Mass
9 characters with decimal point, right justification.
Unit
3 characters, left justification.
Command
3 characters, left justification.
Example 1:
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF - printout generated upon pressing
ENTER/PRINT key.
Example 2:
? _ - _ _ _ _ 2 . 2 3 7 _ l b _ CR LF - printout generated upon pressing
ENTER/PRINT key.
Example 3:
^ _ _ _ _ _ _ 0 . 0 0 0 _ k g _ CR LF - printout generated upon pressing
ENTER/PRINT key.