6. START-UP AND OPERATION
6.1. Levelling
To level the weighing instrument turn its feet. Keep turning the feet until the air
bubble takes central position:
6.2. Connecting the Scale to the Mains
The weighing device can be connected to the mains only with a power supply
that comes standard with the particular model. Nominal voltage of the power
supply (specified on the power supply data plate) has to be compatible with the
mains nominal voltage.
Procedure:
Connect the power supply to the mains. Plug it to the power supply socket
(back side of the scale housing).
Press
key. The key is also used to switch the scale on/off.
Display test proceeds (all symbols are backlit for a moment), program
name and number is displayed first, ZERO indication with reading unit
next (displayed reading unit is conditioned by scale type).
In case the weighing result is not zero after indication stabilisation, press
key.
6.3. Warm-Up Time
For correct scale operation, the workroom temperature must range
between +15°C ÷ +30°C.
It takes 30 minutes for the device to warm up.
During the thermal stabilization, the indications on the screen can change.
If you want to carry out adjustment operation, make sure that your device
is thermally stable.
Temperature and humidity variations during operation may cause
indication errors. To remove the indication errors perform user adjustment.

Table of Contents

Other manuals for RADWAG WTC 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG WTC 2000 and is the answer not in the manual?

RADWAG WTC 2000 Specifications

General IconGeneral
BrandRADWAG
ModelWTC 2000
CategoryScales
LanguageEnglish

Related product manuals