- 45 -
K0
Unlock scale keypad
NB
Give scale serial number
PC
Send all implemented commands
Each command must end with CR LF characters.
27.3. Response Format
On receipt of a command, the indicator responds as follows:
XX_A CR LF
command understood and in progress.
XX_D CR LF
command carried out (appears only after the XX_A command).
XX_I CR LF
command understood but not accessible at this moment.
XX _ ^ CR LF
command understood but max threshold is exceeded.
XX _ v CR LF
command understood but min threshold is exceeded.
ES_CR LF
command not recognised.
XX _ E CR LF
time limit exceeded while waiting for stable measurement result
(time limit is a characteristic scale parameter).
XX - name of a sent command.
_ - space.
27.4. Commands Overview
27.4.1. Zero Scale
Format: Z CR LF
Response options:
Z_A CR LF
Z_D CR LF
- command understood and in progress.
- command carried out.
Z_A CR LF
Z_^ CR LF
- command understood and in progress.
- command understood but zeroing range is exceeded.
Z_A CR LF
Z_E CR LF
- command understood and in progress.
- time limit exceeded while waiting for stable measurement result.
Z_I CR LF
- command understood but not accessible at this moment.