- 46 -
27.4.2. Tare Scale
Format: T CR LF
Response options:
T_A CR LF
T_D CR LF
- command understood and in progress.
- command carried out.
T_A CR LF
T_v CR LF
- command understood and in progress.
- command understood but taring range is exceeded.
T_A CR LF
T_E CR LF
- command understood and in progress.
- time limit exceeded while waiting for stable measurement result.
T_I CR LF
- command understood but not accessible at this moment.
27.4.3. Give Tare Value
Format: OT CR LF
Response: OT_TARE CR LF - command carried out.
Response format:
1 2 3 4 5-6 7-15 16 17 18 19 20 21
O T space
stability
marker
space tare space unit CR LF
Tare - 9 characters, right justification.
Unit - 3 characters, left justification.
27.4.4. Set Tare
Format: UT_TARE CR LF, where TARE - tare value.
Response options:
UT_OK CR LF
- command carried out.
UT_I CR LF
- command understood but not accessible at this moment.
ES CR LF
- command not recognised (tare format incorrect).
Use dot in tare format as decimal point.