- 18 -
Adjustment types:
• external adjustment, <1.1.CA-E>, performed using an external weight of
declared mass, i.e. mass that cannot be modified,
• user adjustment, <1.2.CA-u>, performed using an external weight of mass
of any value comprised within the weighing range, however not lower than
30% of the maximum capacity value.
12.1. External Adjustment
External adjustment must be carried out using an external adjustment weight of
class F
1
.
Procedure:
• Enter <P1.CAL / 1.1.CA-E> submenu, text <UnLoAd> (remove weight) is
displayed.
• Remove the load from the weighing pan and press
key.
• Mass of an empty weighing pan is determined, this is signalled with
display of 'dash', < - >. Next, text <LoAd> (load weight) and mass
value that is to be loaded, e.g. 200 g (conditioned by scale type), are
displayed.
• Load the weighing pan with weight of specified mass value and press
key.
• Weight mass is determined, this is signalled with display of 'dash',
< - >. Next, text <UnLoAd> (remove weight) is displayed.
• Remove the load form the weighing pan, <1.1.CA-E> submenu is
displayed.
12.2. User Adjustment
User adjustment must be carried out using an external adjustment weight of
class F
1
.
Procedure:
• Enter <P1.CAL / 1.2.CA-u> submenu, edit box for declaring weight mass
is displayed (the mass value must be ≥ 30% of the maximum capacity
value).
• Enter weight mass value and press
key to confirm, text <UnLoAd>
(remove weight) is displayed.