- 49 -
27.4.9. Switch on Continuous Transmission in Basic Measuring Unit
Format: C1 CR LF
Response options:
C1_I CR LF
- command understood but not accessible at this moment.
C1_A CR LF
MASS FRAME
- command understood and in progress.
- response: mass value in basic measuring unit.
Response format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space
stability
marker
space character mass space unit CR LF
27.4.10. Switch off Continuous Transmission in Basic Measuring Unit
Format: C0 CR LF
Response options:
C0_I CR LF
- command understood but not accessible at this moment.
C0_A CR LF
- command understood and carried out.
27.4.11. Switch on Continuous Transmission in Current Measuring Unit
Format: CU1 CR LF
Response options:
CU1_I CR LF
- command understood but not accessible at this moment.
CU1_A CR LF
MASS FRAME
- command understood and in progress.
- response: mass value in current measuring unit.
Response format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I
stability
marker
space character mass space unit CR LF
27.4.12. Switch off Continuous Transmission in Current Measuring Unit
Format: CU0 CR LF
Response options: