RAUMED NeuroSmart Instruction Manual
Manufacturer: RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
zwo-541EN Rev 3 / 2022-05-03
Page 36 of 61
Figure 23: Clinic-specific alarm limits
If the alarm limits have been changed, you can restore the clinical alarm limits by pressing the <Load>
button 3B. The default alarm limits shown in can be loaded by pressing the upper Load button <3C>.
Keeping the alarm limits
If you want to keep the specific alarm limits from the last measurement for a new measurement, please
select the alarm limit option “Keep Alarm Limits” when starting the new measurement. See Section 3.4.
Delaying physiological alarms
Delaying high-priority alarms
The alarm condition delay has a maximum value of 15 s + measurement interval, e.g. if the measure-
ment interval is 1 second during oxygen partial pressure measurement, this would result in a maximum
delay to the alarm condition of 16 s (15 s + 1 s).