RAUMED NeuroSmart Instruction Manual
Manufacturer: RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
zwo-541EN Rev 3 / 2022-05-03
Page 56 of 61
Invasive pressure
Number of channels:
2
Measuring position designation:
ICP1, ICP2, ART, CVP
Resolution:
0.1 mmHg
Measurement range:
- 40 ... + 400 mmHg
Offset adjustment range:
- 150 ... + 150 mmHg
Accuracy:
1 mmHg for the measuring equipment and 1 mmHg for the
catheter
Sampling rate:
100 Hz
Defibrillator protection
Yes
Pressure sensor
Supply voltage:
4.1 V
Sensor input impedance
> 1 kΩ
Sensor sensitivity:
5 V / V / mmHg
Telemetry channel pressure
Number of channels:
1
Measuring position designation:
ICPT
Resolution:
1 mmHg
Measurement range:
- 40 ... + 400 mmHg
Accuracy:
±2 mmHg
Sampling rate:
5 Hz
Defibrillator protection
Yes
Oxygen partial pressure
invasive
(only applicable to REF 095294-
001)
Number of channels:
1
Measuring position designation:
pO2
Resolution:
0.1 mmHg
Measurement range:
0 ... + 200 mmHg
Accuracy:
See accuracy of oxygen sensor
Sampling rate:
1 Hz
Defibrillator protection
Yes
Oxygen sensor
Fibre optic
Light source
Green LED