Setting Alarm and Recording Limits
24
Menu Option LCD Display Description of Option or Instructions Menu Path
ECG Record RECORD ECG?
This parameter allows the user to select whether
or not the ECG waveform will be recorded.
Values:
YES or NO
Standard value: YES
All Menus
Record Menu
Auxiliary 1
Record
RECORD
AUXILIARY 1
This parameter allows the user to select whether
or not waveforms will be recorded on Auxiliary 1.
Standard value: OFF
For additional information refer to the Auxiliary
Signal Interface section of this manual.
All Menus
Record Menu
Auxiliary 2
Record
RECORD
AUXILIARY 2
This parameter allows the user to select whether
or not waveforms will be recorded on Auxiliary 2.
Standard value: OFF
For additional information refer to the Auxiliary
Signal Interface section of this manual.
All Menus
Record Menu
Auxiliary 3
Record
RECORD
AUXILIARY 3
This parameter allows the user to select whether
or not waveforms will be recorded on Auxiliary 3.
Standard value: OFF
For additional information refer to the Auxiliary
Signal Interface section of this manual.
All Menus
Record Menu
Auxiliary 4
Record
RECORD
AUXILIARY 4
This parameter allows the user to select whether
or not waveforms will be recorded on Auxiliary 4.
Standard value: OFF
For additional information refer to the Auxiliary
Signal Interface section of this manual.
All Menus
Record Menu
External
Physiological
Trigger
EXT. PHYSIO
TRIGGER
Ext. Physio. Trigger allows external equipment to
trigger a SmartMonitor 2 recording when a
physiological parameter is violated in the external
auxiliary device.
Select from:
OFF
Trigger when high
Trigger when low
See Using Auxiliary Equipment
All Menus
Record Menu
External
Equipment
Trigger
EXT. EQUIP.
TRIGGER
Ext. Equip. Trigger allows external equipment to
cause an entry in the SmartMonitor 2 Equipment
Log when an equipment parameter is violated in
the external auxiliary device.
Select from:
OFF
Trigger when high
Trigger when low
See Using Auxiliary Equipment
All Menus
Record Menu