Chapter 6 Operations and Measurements RIGOL
MSO4000/DS4000 User’s Guide 6-47
cursor A and cursor B positions are defined to “0°” and “360°” phase
positions respectively; at the same time, two unmovable vertical cursor
lines (with relatively lower brightness) are displayed on the screen as the
phase reference positions (“0°” and “360°” phase positions are marked by a
solid line and dotted line respectively). Before you setting them manually,
the oscilloscope will use the default phase reference positions.
%: measure the ratio values at X cursors. In the measurement results, CurA,
CurB and X are all expressed in percentage.
The oscilloscope allows you to set the ratio reference positions according to
your need. After adjusting the X cursors to the desired positions (refer to
step 4 Ajust the X cursor positions”), press SetCursor and the current
cursor A and cursor B positions are defined to “0%” and “100%” ratio
positions respectively; at the same time, two unmovable vertical cursor
lines (with relatively lower brightness) are displayed on the screen as the
ratio reference positions (“0%” and “100%” ratio positions are marked by a
solid line and dotted line respectively). Before you setting them manually,
the oscilloscope will use the default ratio reference positions.
4. Ajust the X cursor positions
1) Press CursorA and turn the multi-function knob to adjust the
horizontal position of cursor A (X cursor). The adjustable range is limited
within the screen.
2) Press CursorB and turn the multi-function knob to adjust the
horizontal position of cursor B (X cursor). The adjustable range is limited
within the screen.
3) Press CursorAB and turn the multi-function knob to adjust the
horizontal positions of cursors A and B (X cursors) at the same time. The
adjustable range is limited within the screen. The horizontal distance
between cursor A and B (X cursors) keeps unchanged.
Note: When any of CursorA, CursorB and CursorAB is currently selected,
you can press the multi-function knob
continuously to quickly switch among
the three items.
www.calcert.com sales@calcert.com1.800.544.2843
0
5
10
15
20
25
30

Table of Contents

Other manuals for Rigol DS4024

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS4024 and is the answer not in the manual?

Rigol DS4024 Specifications

General IconGeneral
BrandRigol
ModelDS4024
CategoryTest Equipment
LanguageEnglish

Related product manuals