Status Description Bit
001 I/O 1 state 0
002 I/O 2 state 1
004 I/O 3 state 2
008 I/O 4 state 3
016 I/O 5 state 4
032 I/O 6 state 5
064 I/O 7 state 6
128 I/O 8 state 7
Example
Command Answer Description
S01; Select unit 1
COF?; 3 CRLF Query format
MSV?; -00001.0 CRLF Query weight reading
COF9; 0 CRLF Change to format 9
TDD1; 0 CRLF Save new setting
MSV?; -00001.0,01,006 CRLF Query weight reading using the new format
21.3.14 CWT: Calibration Weight
Set the calibration weight to be used for span calibration. This must be set before using the LWT; or LWN;
span calibration commands.
General
No. of parameters 1
Save changes With TDD1
Increment Trade Counter No
Parameter Details
Parameter Description Range Default
1 Calibration Weight 2% - 100% of full scale weight. Send IAD? to read full scale
setting.
3000
Example
Command Answer Description
S01; Select unit 1
CWT?; 3000 CRLF Query calibration weight setting
CWT4000; 0 CRLF Change calibration weight to 4000. Note that weight is sent without any decimal
point, so 400.0 kg is sent as 4000 not 400.0.
TDD1; 0 CRLF Save new setting
21.3.15 DFT: Default Values
Set the user database, runtime database, calibration, Ethernet settings or licensing database to defaults.
General
No. of parameters 1
Save changes At input
Increment Trade Counter Yes (only Calibration)
Parameter Details
130 C500-600-2.0.3