Range 1 verification in-
terval
⊕ E1 Sets the minimum verification interval of the scale (as dis-
played). If BUILD:TYPE=DUAL I or BUILD:TYPE=DUAL R, it sets
minimum verification interval of the 1st range. Options are:
1(def)
2
5
10
20
50
100
Range 2 scale capacity ⊕ CAP2 Sets the maximum capacity of range 2. Only used if BUILD:
TYPE=DUAL I or BUILD:TYPE=DUAL R. Default: 6000.
Range 2 verification in-
terval
⊕ E2 Sets the minimum verification interval of range 2. Only used if
BUILD:TYPE=DUAL I or BUILD:TYPE=DUAL R. Options are:
1(def)
2
5
10
20
50
100
Additive tare limit ⊕ AD.TARE Sets the additive tare limit. Default: 0. See Section 7.2.6 for
more details.
Weighed units ⊕ UNITS Sets the weighing units for the scale. Options are:
NONE
G
KG (default)
LB
T
OZ
USER
A/D synchronisation
frequency (effects
calibration)
⊕ SYNC Sets the frequency of the primary anti-noise filter in Hz. This
sets the rate that data is processed. It will deliver optimum
noise performance in a 50Hz environment when set to 12.5, 25
or 50. Options are: 10, 12.5, 15, 20, 25, 30, 50(def), 60, 100
High resolution display ⊕ HI.RES In full setup, this setting enables x10 high resolution display.
Weights will be shown with 10x resolution. The options are:
OFF(def)
ON
In safe setup, this setting will display the weight in x10 mode
only while the menu is active. Upon exit the x10 mode is dis-
abled. In trade mode this test is only available for 5 seconds.
⊕ = Change only possible in FULL Setup
19 C500-600-2.0.3