161
Serial Interface
Measurement Time Manual (Num)
Measurement time of user setting on manual store (number)
Setting value when “Measurement Time Preset” command parameter
is “Manual” on manual store mode
Setting command Measurement
Time Manual (Num), p1
Parameter p1= 1 to 59 (Time unit is s [second] or m [minute])
p1= 1 to 24 (Time unit is h [hour])
Request command Measurement
Time Manual (Num)?
Response data d1
Returned value Same as for setting command
Measurement Time Manual (Unit)
Measurement time of user setting on manual store (unit)
Setting time unit when “Measurement Time Preset” command param-
eter is “Manual” on manual store mode
Setting command Measurement Time Manual (Unit), p1
Parameter p1= “s”
p1= “m”
p1= “h”
Request command Measurement
Time Manual (Unit)?
Response data d1
Returned value Same as for setting command