Cedex Bio System 9 Messages and alarms
List of alarm messages
7038 [%s] Motor %d error (overheated). Switch off the instrument. Check for smooth transfer
movements. Restart the system. If the problem persists
call service.
7039 [%s] Power supply error. One or more internally used
voltages are outside the range.
Switch off the instrument. Check the fuses F1 through F4
and replace blown fuses. Restart the system. If the error
persists call service.
7042 [%s] Fuse F1 blown. Switch off the instrument. Check fuse F1. If necessary
replace fuse F1. Restart the system. If the error persists call
service.
7043 [%s] Fuse F2 blown. Switch off the instrument. Check fuse F2. If necessary
replace fuse F2. Restart the system. If the error persists call
service.
7044 [%s] Fuse F3 blown. Switch off the instrument. Check fuse F3. If necessary
replace fuse F3. Restart the system. If the error persists call
service.
7045 [%s] Fuse F4 blown. Switch off the instrument. Check fuse F4. If necessary
replace fuse F4. Restart the system. If the error persists call
service.
7046 [%s] The initialization of the rotor failed. Switch off the instrument. Check if the rotor is jamming
or blocked. Clean the rotor initialization light barrier.
Restart the system. If the error persists call service.
7051 [%s] Error rotor motor. The motor could not be started. Restart the system. If the error persists call service.
7052 [%s] Error motor X-transfer. The motor could not be
started.
Restart the system. If the error persists call service.
7053 [%s] Error motor Z-transfer. The motor could not be
started.
Restart the system. If the error persists call service.
7057 [%s] The ISE sample sensor is not calibrated or no
calibration data is available.
Perform <Initialize ISE Unit>. If the error persists call ser-
vice.
7058 [%s] ISE sample sensor could not detect fluid. Check the ISE tower and the ISE tubing for obstructions
and leaks. Ensure that the electrodes are installed cor-
rectly. If the error persists call service.
7059 [%s] The ISE sample sensor could not detect the end of
the transported fluid segment.
Check the ISE tower and the ISE tubing for obstructions
and leaks. If the error persists call service.
7060 [%s] The calibration of the ISE sample sensor failed. Check the filling level of the ISE Cal bottle. Check the ISE
tubing for obstructions and leaks. Perform <Initialize ISE
Unit>. If the error persists call service.
7061 [%s] The calibration of the ISE sample sensor failed. Check the filling level of the ISE Cal bottle. Check the ISE
tubing for obstructions and leaks. Perform <Initialize ISE
Unit>. If the error persists call service.
7062 [%s] The ISE sample sensor could not detect calibrator. Check the filling level of the ISE Cal bottle. Check the ISE
tubing for obstructions and leaks. Perform <Initialize ISE
Unit>. If the error persists call service.
7063 [%s] The ISE sample sensor could not detect the end of
the calibrator segment.
Check the ISE tubing for obstructions and leaks. Perform
<Initialize ISE Unit>. If the error persists call service.
7064 [%s] The ISE sample sensor could not detect Activator
during initialization.
Check the ISE tower and ISE tubing for obstructions and
leaks. Perform <Initialize ISE Unit>. If the error persists
call service.
7065 [%s] Calibration of the ISE reference sensor failed, no
valid calibration data are available.
Check the filling level of ISE Ref. Perform <Initialize ISE
Reference Sensor>. If the error persists call service.
ID Message Comment
Table D-1 List of alarm messages (Continued)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals