Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
189User Manual 1173.9240.02 ─ 17
[SENSe:]MIXer:HARMonic:HIGH:STATe <State>
This command specifies whether a second (high) harmonic is to be used to cover the
band's frequency range.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
MIX:HARM:HIGH:STAT ON
Manual operation: See "Range 1/2" on page 61
[SENSe:]MIXer:HARMonic:HIGH[:VALue] <HarmOrder>
This command specifies the harmonic order to be used for the high (second) range.
Parameters:
<HarmOrder numeric value
Range: 2 to 61 (USER band); for other bands: see band
definition
Example:
MIX:HARM:HIGH 2
Manual operation: See "Harmonic Order" on page 61
[SENSe:]MIXer:HARMonic:TYPE <OddEven>
This command specifies whether the harmonic order to be used should be odd, even,
or both.
Which harmonics are supported depends on the mixer type.
Parameters:
<OddEven> ODD | EVEN | EODD
*RST: EVEN
Example:
MIX:HARM:TYPE ODD
Manual operation: See "Harmonic Type" on page 61
[SENSe:]MIXer:HARMonic[:LOW] <HarmOrder>
This command specifies the harmonic order to be used for the low (first) range.
Parameters:
<HarmOrder> numeric value
Range: 2 to 61 (USER band); for other bands: see band
definition
*RST: 2 (for band F)
Example:
MIX:HARM 3
Manual operation: See "Harmonic Order" on page 61
Configuring the Measurement