Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
188User Manual 1173.9240.02 ─ 17
[SENSe:]MIXer:HARMonic:BAND:PRESet
This command restores the preset frequency ranges for the selected standard wave-
guide band.
Note: Changes to the band and mixer settings are maintained even after using the
PRESET function. Use this command to restore the predefined band ranges.
Example:
MIX:HARM:BAND:PRES
Presets the selected waveguide band.
Usage: Event
Manual operation: See "Preset Band" on page 60
[SENSe:]MIXer:HARMonic:BAND[:VALue] <Band>
This command selects the external mixer band. The query returns the currently
selected band.
This command is only available if the external mixer is active (see [SENSe:]MIXer[:
STATe] on page 185).
Parameters:
<Band> KA | Q | U | V | E | W | F | D | G | Y | J | USER
Standard waveguide band or user-defined band.
Manual operation: See "Band" on page 60
Table 11-2: Frequency ranges for pre-defined bands
Band Frequency start [GHz] Frequency stop [GHz]
KA (A) *) 26.5 40.0
Q 33.0 50.0
U 40.0 60.0
V 50.0 75.0
E 60.0 90.0
W 75.0 110.0
F 90.0 140.0
D 110.0 170.0
G 140.0 220.0
J 220.0 325.0
Y 325.0 500.0
USER 32.18
(default)
68.22
(default)
*) The band formerly referred to as "A" is now named "KA".