Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
185User Manual 1173.9240.02 ─ 17
[SENSe:]MIXer:LOPower................................................................................................185
[SENSe:]MIXer:SIGNal...................................................................................................186
[SENSe:]MIXer:THReshold............................................................................................. 186
[SENSe:]MIXer[:STATe] <State>
Activates or deactivates the use of a connected external mixer as input for the mea-
surement. This command is only available if the R&S FSW-B21 option is installed and
an external mixer is connected.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
MIX ON
Manual operation: See "External Mixer State" on page 59
[SENSe:]MIXer:BIAS:HIGH <BiasSetting>
This command defines the bias current for the high (second) range.
This command is only available if the external mixer is active (see [SENSe:]MIXer[:
STATe] on page 185).
Parameters:
<BiasSetting> *RST: 0.0 A
Default unit: A
Manual operation: See "Bias Settings" on page 63
[SENSe:]MIXer:BIAS[:LOW] <BiasSetting>
This command defines the bias current for the low (first) range.
This command is only available if the external mixer is active (see [SENSe:]MIXer[:
STATe] on page 185).
Parameters:
<BiasSetting> *RST: 0.0 A
Default unit: A
Manual operation: See "Bias Settings" on page 63
[SENSe:]MIXer:LOPower <Level>
This command specifies the LO level of the external mixer's LO port.
Parameters:
<Level> numeric value
Range: 13.0 dBm to 17.0 dBm
Increment: 0.1 dB
*RST: 15.5 dBm
Configuring the Measurement