Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
208User Manual 1173.9240.02 ─ 17
Return values:
<Name> Name string
Usage: Query only
[SENSe:]PROBe<p>:SETup:STATe?
Queries if the probe at the specified connector is active (detected) or not active (not
detected). To switch the probe on, i.e. activate input from the connector, use
INP:SEL:AIQ (see INPut:SELect on page 184).
Suffix:
<p>
.
1 | 2 | 3
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF (currently not supported; use "1" with RF Input Connec-
tor setting "Baseband Input I")
Return values:
<State> DETected | NDETected
*RST: NDETected
Usage: Query only
[SENSe:]PROBe<p>:SETup:TYPE?
Queries the type of the probe.
Suffix:
<p>
.
1 | 2 | 3
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF (currently not supported; use "1" with RF Input Connec-
tor setting "Baseband Input I")
Return values:
<Type> String containing one of the following values:
– None (no probe detected)
– active differential
– active single-ended
Usage: Query only
11.4.2.6 Working with Power Sensors
The following commands describe how to work with power sensors.
● Configuring Power Sensors.................................................................................. 209
● Configuring Power Sensor Measurements........................................................... 210
● Triggering with Power Sensors............................................................................. 217
Configuring the Measurement