Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
218User Manual 1173.9240.02 ─ 17
[SENSe:]PMETer<p>:TRIGger:HYSTeresis <Hysteresis>
This command defines the trigger hysteresis for external power triggers.
The hysteresis in dB is the value the input signal must stay below the IF power trigger
level in order to allow a trigger to start the measurement.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Hysteresis> Range: 3 dB to 50 dB
Increment: 1 dB
*RST: 0 dB
Example:
PMET2:TRIG:HYST 10
Sets the hysteresis of the trigger to 10 dB.
Manual operation: See "Hysteresis" on page 85
[SENSe:]PMETer<p>:TRIGger:LEVel <Level>
This command defines the trigger level for external power triggers.
This command requires the use of an R&S NRP-Z81 power sensor.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Level> -20 to +20 dBm
Range: -20 dBm to 20 dBm
*RST: -10 dBm
Example:
PMET2:TRIG:LEV -10 dBm
Sets the level of the trigger
Manual operation: See "External Trigger Level" on page 84
[SENSe:]PMETer<p>:TRIGger:SLOPe <Edge>
This command selects the trigger condition for external power triggers.
Suffix:
<p>
.
1...4
Power sensor index
Configuring the Measurement