Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
219User Manual 1173.9240.02 ─ 17
Parameters:
<Edge> POSitive
The measurement starts in case the trigger signal shows a posi-
tive edge.
NEGative
The measurement starts in case the trigger signal shows a neg-
ative edge.
*RST: POSitive
Example:
PMET2:TRIG:SLOP NEG
Manual operation: See "Slope" on page 85
[SENSe:]PMETer<p>:TRIGger[:STATe] <State>
This command turns the external power trigger on and off.
This command requires the use of an R&S NRP-Z81 power sensor.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON | OFF
*RST: OFF
Example:
PMET2:TRIG ON
Switches the external power trigger on
Manual operation: See "Using the power sensor as an external trigger" on page 84
11.4.2.7 External Generator Control
External generator control commands are available if the R&S FSW External Genera-
tor Control option (R&S FSW-B10) is installed. For each measurement channel one
external generator can be configured. To switch between different configurations
define multiple measurement channels.
For more information on external generator control see chapter 4.7.4, "Basics on Exter-
nal Generator Control", on page 36.
● Measurement Configuration..................................................................................219
● Interface Configuration..........................................................................................223
● Source Calibration.................................................................................................225
● Programming Example for External Generator Control........................................ 227
Measurement Configuration
The following commands are required to activate external generator control and to con-
figure a calibration measurement with an external tracking generator.
Configuring the Measurement