Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
223User Manual 1173.9240.02 ─ 17
Interface Configuration
The following commands are required to configure the interface for the connection to
the external generator.
SOURce:EXTernal:ROSCillator[:SOURce]........................................................................223
SYSTem:COMMunicate:GPIB:RDEVice:GENerator:ADDRess............................................223
SYSTem:COMMunicate:RDEVice:GENerator:INTerface.................................................... 223
SYSTem:COMMunicate:RDEVice:GENerator:LINK........................................................... 224
SYSTem:COMMunicate:RDEVice:GENerator:TYPE..........................................................224
SYSTem:COMMunicate:TCPip:RDEVice:GENerator:ADDRess.......................................... 224
SOURce:EXTernal:ROSCillator[:SOURce] <Source>
This command controls selection of the reference oscillator for the external generator.
If the external reference oscillator is selected, the reference signal must be connected
to the rear panel of the instrument.
Parameters:
<Source> INTernal
the internal reference is used
EXTernal
the external reference is used; if none is available, an error flag
is displayed in the status bar
*RST: INT
Example:
SOUR:EXT:ROSC EXT
Switches to external reference oscillator
Manual operation: See "Reference" on page 75
SYSTem:COMMunicate:GPIB:RDEVice:GENerator:ADDRess <Number>
Changes the IEC/IEEE-bus address of the external generator.
Parameters:
<Number> Range: 0 to 30
*RST: 28
Example:
SYST:COMM:GPIB:RDEV:GEN:ADDR 15
Manual operation: See "GPIB Address / TCP/IP Address" on page 75
SYSTem:COMMunicate:RDEVice:GENerator:INTerface <Type>
Defines the interface used for the connection to the external generator.
This command is only available if external generator control is active (see SOURce:
EXTernal[:STATe] on page 222).
Parameters:
<Type> GPIB | TCPip
*RST: GPIB