Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
250User Manual 1173.9240.02 ─ 17
TRIGger[:SEQuence]:SLOPe <Type>
For all trigger sources except time you can define whether triggering occurs when the
signal rises to the trigger level or falls down to it.
For external and time domain trigger sources you can define whether triggering occurs
when the signal rises to the trigger level or falls down to it.
Parameters:
<Type> POSitive | NEGative
POSitive
Triggers when the signal rises to the trigger level (rising edge).
NEGative
Triggers when the signal drops to the trigger level (falling edge).
*RST: POSitive
Example:
TRIG:SLOP NEG
Manual operation: See "Slope" on page 99
TRIGger[:SEQuence]:SOURce <Source>
This command selects the trigger source.
For triggering with AF, AM, AMRelative, FM, and PM trigger sources to be successful,
the measurement time must cover at least 5 periods of the audio signal. For details on
trigger sources see "Trigger Source" on page 95.
Note on external triggers:
If a measurement is configured to wait for an external trigger signal in a remote control
program, remote control is blocked until the trigger is received and the program can
continue. Make sure this situation is avoided in your remote control programs.