Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
249User Manual 1173.9240.02 ─ 17
Manual operation: See "Trigger Level" on page 98
TRIGger[:SEQuence]:LEVel:AM[:ABSolute] <Level>
The command sets the level when RF power signals are used as trigger source.
For triggering to be successful, the measurement time must cover at least 5 periods of
the audio signal.
Parameters:
<Level> Range: -100 to +30
*RST: -20 dBm
Default unit: dBm
Example:
TRIG:LEV:AM -30 dBm
Sets the RF power signal trigger threshold to -30 dBm
Manual operation: See "Trigger Level" on page 98
TRIGger[:SEQuence]:LEVel:FM <Level>
The command sets the level when FM-modulated signals are used as trigger source.
For triggering to be successful, the measurement time must cover at least 5 periods of
the audio signal.
Parameters:
<Level> Range: -10 to +10
*RST: 0 Hz
Default unit: MHz
Example:
TRIG:LEV:FM 10 kHz
Sets the FM trigger threshold to 10 kHz
Manual operation: See "Trigger Level" on page 98
TRIGger[:SEQuence]:LEVel:PM <Level>
The command sets the level when PM-modulated signals are used as trigger source.
For triggering to be successful, the measurement time must cover at least 5 periods of
the audio signal.
Parameters:
<Level> Range: -1000 to +1000
*RST: 0 RAD
Default unit: RAD | DEG
Example:
TRIG:LEV:PM 1.2 RAD
Sets the PM trigger threshold to 1.2 rad
Manual operation: See "Trigger Level" on page 98